2Ć126
MEASUrement:REFLevel:PERCent:HIGH
Sets or queries the percent, relative to HIGH, that is used to calculate the high
reference level when MEASUrement:REFLevel:METHod is set to PERCent.
This command is equivalent to setting the Reference Levels in the Measure
menu.
Measurement
MEASUrement:REFLevel:PERCent:HIGH <NR3>
MEASUrement:REFLevel:PERCent:HIGH?
HIGH
?
<Space> <NR3>
MEASUrement :
REFLevel
:
PERCent
:
<NR3> ranges from 0 to 100%, and is the high reference level. The default is
90%.
MEASUREMENT:REFLEVEL:PERCENT:HIGH 95
specifies that the high reference level is set to 95% of HIGH.