Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć31
ACQuire:NUMENv
Sets the number of waveform acquisitions that make up an envelope waveĆ
form. This is equivalent to setting the Envelope count in the Acquisition Mode
side menu.
Acquisition
ACQuire:MODe
ACQuire:NUMENv { <NR1> | INFInite }
ACQuire:NUMENv?
ACQuire
:
<NR1>
?
NUMENv
<Space
>
INFInite
<NR1> 0 is the number of waveform acquisitions, from 2 to 256. The enveĆ
lope will restart after the specified number of envelopes have been acquired
or when the ACQuire:STATE RUN command is sent.
INFInite or <NR1> = 0 specifies continuous enveloping.
NOTE
If you set the acquisition system to single sequence envelope mode
and set the number of envelopes to infinity, the oscilloscope will
envelope a maximum of 257 acquisitions.
ACQUIRE:NUMENV 10
specifies that an enveloped waveform will show the result of combining
10 separately acquired waveforms.
ACQUIRE:NUMENV?
might return 0, indicating that acquisitions are acquired infinitely for
enveloped waveforms.
Group:
Related Commands:
Syntax:
Arguments:
Examples: