Tektronix TDS 320 Sprinkler User Manual


 
Command Descriptions
Syntax and Commands
2Ć132
*PUD
Sets or queries a string of Protected User Data. This data is protected by the
PASSWord command. You can modify it only by first entering the correct
password. The password is not necessary to query the data.
Miscellaneous
PASSWord
*PUD <Block>
*PUD?
*PUD
?
<Space>
<Block>
<Block> is a string containing up to 100 characters.
*PUD #229This instrument belongs to me
stores the string This instrument belongs to me" in the user protected
data area.
*PUD?
might return #221Property of Company X.
*RCL (No Query Form)
Restores the state of the oscilloscope from a copy of its settings stored in
memory. (The settings are stored using the *SAV command.) This command
is equivalent to RECAll:SETUp, and performs the same function as the Recall
Saved Setup item in the frontĆpanel Save/Recall Setup menu.
Save and Recall
FACtory, *LRN?, RECAll:SETUp, *RST, *SAV, SAVe:SETUp
*RCL <NR1>
<NR1>*RCL <Space>
<NR1> is a value in the range from 1 to 10, and specifies a setup storage
location. Using an outĆofĆrange value causes an execution error (222, Data
out of range").
*RCL 3
restores the oscilloscope from a copy of the settings stored in memory
location 3.
Group:
Related Commands:
Syntax:
Arguments:
Examples:
Group:
Related Commands:
Syntax:
Arguments:
Examples: