Tektronix TDS 320 Sprinkler User Manual


 
Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć81
DISplay:PERSistence
Sets the length of time that dots (or vectors) are displayed when DISĆ
play:STYle is set to ACCUMDOTS or ACCUMVECTORS.
Display
DISplay:STYle
DISplay:PERSistence { INFinite | <NR3> }
DISplay:PERSistence?
INFinite
DISplay
:
PERSistence
?
<Space>
<NR3>
<NR3> specifies the length, in seconds, that the waveform points are disĆ
played on the screen. The range is 500 ms to 10 s.
INFinite specifies infinite persistence.
DISPLAY:PERSISTENCE 3
specifies that the waveform points are displayed on the screen for 3 seĆ
conds before they fade.
Group:
Related Commands:
Syntax:
Arguments:
Examples: