Tektronix TDS 320 Sprinkler User Manual


 
Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć151
TRIGger:MAIn
Sets the main trigger level and returns the current main trigger parameters.
Trigger
TRIGger:MAIn SETLevel
TRIGger:MAIn?
TRIGger :
MAIn
SETLevel
?
<Space>
SETLevel sets the main trigger level to half way between the MIN and MAX
amplitudes of the trigger source input. This is equivalent to pressing the
frontĆpanel SET LEVEL TO 50% button.
TRIGGER:MAIN SETLEVEL
sets the main trigger level mid way between MAX and MIN.
TRIGger:MAIn:EDGE? (Query Only)
Returns the trigger coupling, source, and slope for the main edge trigger.
Trigger
TRIGger:MAIn:EDGE?
:
MAIn
TRIGger
EDGE ?:
TRIGGER:MAIN:EDGE?
might return SOURCE CH1;COUPLING DC;SLOPE RISE
Group:
Syntax:
Arguments:
Examples:
Group:
Syntax:
Examples: