Agilent Technologies E4371A Lawn Mower User Manual


 
Language Dictionary - 6
71
cfCalTransfer
CAUTION: Make sure that no cells are connected when executing cfCalTransfer.
Syntax
int cfCalTransfer(CF_HANDLE server);
Description
Begins a transfer calibration sequence. This function uses the instrument’s internal references to
calibrate the measurement and output circuits of each channel. There should not be any loads or
cells connected to the outputs when this command is given.
Since transfer calibration can take up to 15 minutes, calibration functions do not wait for
calibration to complete. They return immediately after starting calibration. To monitor the
progress and results of calibration, use cfGetInstStatus. While calibration is in progress, the
CF_CALIBRATING_STAT bit is true. If any errors occur during calibration, the
CF_CAL_ERROR_STAT bit is true. Details of the errors can be obtained using cfReadTestLog.
cfClose
Syntax
int cfClose(CF_HANDLE server);
Description
Closes a server connection. An Agilent MCCD server can only accommodate a limited number
of open client connections. Closing a connection makes it available to other clients. Closing a
connection does not affect the output functions of the server, and all other instrument functions
continue to operate undisturbed by cfOpen and cfClose commands.
cfDeleteGroup
Syntax
int cfDeleteGroup(CF_HANDLE server);
Description
Deletes a group from the Agilent MCCD. Server is a handle that was previously obtained by a
call to cfOpenGroup().