Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć67
DATa:STARt
Sets or queries the starting data point for waveform transfer. This command
allows for the transfer of partial waveforms to and from the oscilloscope.
Waveform
CURVe?, DATa SNAp, DATa:STOP
DATa:STARt <NR1>
DATa:STARt?
DATa
:
STARt
?
<Space> <NR1>
<NR1> ranges from 1 to the record length and is the first data point that will
be transferred. Data is transferred from <NR1> to DATa:STOP or the record
length, whichever is less. If <NR1> is greater than the record length, then the
oscilloscope transfers data until it reaches the record length. When
DATa:STOP is less than DATa:STARt, the values are swapped internally for
CURVe?.
DATA:START 10
specifies that the waveform transfer will begin with data point 10.
DATA:START?
might return 214 as the first waveform data point that will be transferred.
Group:
Related Commands:
Syntax:
Arguments:
Examples: