Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć131
*PSC
Sets and queries the powerĆon status flag that controls the automatic powerĆ
on handling of the DESER, SRER, and ESER registers. When *PSC is true,
the DESER register is set to 255 and the SRER and ESER registers are set to
0 at power on. When *PSC is false, the current values in the DESER, SRER,
and ESER registers are preserved in nonvolatile memory when power is shut
off and are restored at power on. For a complete discussion of the use of
these registers, see page 3Ć1.
Status and Error
DESE, *ESE, FACtory, *RST, *SRE
*PSC <NR1>
*PSC?
*PSC
?
<Space>
<NR1>
<NR1> = 0 sets the powerĆon status clear flag to false, disables the power on
clear, and allows the oscilloscope to possibly assert SRQ after power on.
<NR1> 0 sets the powerĆon status clear flag true. Sending *PSC 1, thereĆ
fore, enables the powerĆon status clear and prevents any SRQ assertion after
powerĆon. Using an outĆofĆrange value causes an execution warning.
*PSC 0
sets the powerĆon status clear flag to false.
*PSC?
might return the value 1, showing that the powerĆon status clear flag is set
to true.
Group:
Related Commands:
Syntax:
Arguments:
Examples: