Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć135
RS232?
might return: RS232:BAUD 9600;MODE RAW;PACE XON;PARITY
NONE;SBITS 2;CONTROL:RTS RFR;RS232:TRANSMIT;TERMINATOR
LF;DELAY 0.0
RS232:BAUd
Sets or queries the RSĆ232 interface transmission speed. If no flow control
(flagging) is used, commands may be received faster than the oscilloscope
can process them. Also, if another command is sent immediately after this
command, without waiting for the baud rate to be programmed, the first
couple of characters may be lost.
Miscellaneous
RS232: HARDFLAGGING, RS232: PARITY, RS232:SOFTFLAGGING, RS232:
STOPBITS, RS232?
RS232:BAUd <NR1>
RS232:BAUd?
BAUd
:
<Space>
?
RS232
<NR1>
<NR1> where <NR1> can be 300, 600, 1200, 2400, 4800, 9600 or 19200.
RS232:BAUD 9600
sets the transmission rate to 9600 baud.
RS232:CONTrol:RTS
Sets or queries the state of RS232 hard flagging. This command performs the
same function as RS232:HARDFlagging.
Miscellaneous
RS232: HARDFLAGGING, RS232:SOFTFLAGGING, RS232: STOPBITS,
RS232?
RS232:CONTrol:RTS { ON | RFR | IBFull }
RS232:CONTrol:RTS?
Examples:
Group:
Related Commands:
Syntax:
Arguments:
Examples:
Group:
Related Commands:
Syntax: