Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć49
CH<x>:VOLts
Sets or queries the vertical gain of the specified channel. This command is
identical to the CH<x>:SCAle command and is included for compatibility
purposes. Only CH<x>:SCAle is returned in response to a CH<x>? query.
Vertical
CH1:SCAle
CH<x>:VOLts <NR3>
CH<x>:VOLts?
CH
<x> :
VOLts
?
<Space> <NR3>
CH1:VOLTS 100E–03
sets the channel 1 gain to 100ĂmV/div.
CH2:VOLTS?
might return 1.00E+0, indicating that the current V/div setting of channel
2 is 1ĂV/div.
CLEARMenu (No Query Form)
Clears the current menu from the display. This command is equivalent to
pressing the CLEAR MENU button on the front panel.
Display
CLEARMenu
CLEARMenu
CLEARMENU
clears the menu from the display.
Group:
Related Commands:
Syntax:
Examples:
Group:
Syntax: