Tektronix TDS 310 Sprinkler User Manual


 
Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć85
*ESR? (Query Only)
Returns the contents of the Standard Event Status Register (SESR). *ESR?
also clears the SESR (since reading the SESR clears it). For a more detailed
discussion of the use of these registers see page 3Ć1.
Status and Error
ALLEv?, *CLS, DESE, *ESE, EVENT?, EVMsg?, *SRE, *STB?
*ESR?
?
*ESR
*ESR?
might return the value 213, showing that the SESR contains binary
11010101.
EVENT? (Query Only)
Returns from the Event Queue an event code that provides information about
the results of the last *ESR? read. EVENT? also removes the returned value
from the Event Queue. A discussion of event handling begins on pageĂ3Ć1.
Status and Error
ALLEv?, *CLS, DESE, *ESE, *ESR?, EVMsg?, *SRE, *STB?
EVENT?
?
EVENT
EVENT?
might return the response :EVENT 110, showing that there was an error
in a command header.