Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć147
SET? (Query Only)
Returns a string listing the oscilloscope settings, except for configuration
information for the calibration values. You can use this string to return the
oscilloscope to the state it was in when you sent SET?. This command is
identical to the *LRN? command.
Miscellaneous
HEADer, *LRN?, VERBose
SET?
SET
?
NOTE
The SET? query always returns a string with command headers,
regardless of the setting of the HEADer command. This is because
the returned string is intended to be able to be sent back to the
oscilloscope as a command string. The VERBose command can still
be used to specify whether the returned headers should be abbreĆ
viated or full length.
SET?
a partial return string may look like this:
:ACQUIRE:STOPAFTER RUNSTOP;STATE 1;MODE SAMPLE;NUMENV
8;NUMAVG 16;:HEADER 1;:VERBOSE 1; :ALIAS:STATE 0;:DIS-
PLAY:FORMAT YT;STYLE VECTORS;PERSISTENCE
500.0E–3;GRATICULE FULL;TRIGT 1;INTENSITY:OVERALL
85;WAVEFORM BRIGHT;TEXT DIM;CONTRAST 150;:LOCK
NONE;:HARDCOPY:FORMAT EPSIMAGE;PORT GPIB;LAYOUT POR-
TRAIT;
Group:
Related Commands:
Syntax:
Examples: