Application Examples
TDS1000/2000-Series Digital Oscilloscope User Manual
59
To set up to measure propagation delay, follow these steps:
1. If the channels are not displayed, push the CH 1 MENU and then
CH 2 MENU buttons.
2. Push the AUTOSET button to trigger a stable display.
3. Adjust the horizontal and vertical controls to optimize the
display.
4. Push the CURSOR button to see the Cursor Menu.
5. Push the Type option button and select Time.
6. Push the Source option button and select CH1.
7. Turn the CURSOR 1 knob to place the cursor on the active edge
of the chip-select signal.
8. Turn the CURSOR 2 knob to place the second cursor on the data
output transition.
9. Read the propagation delay in the Delta readout in the Cursor
Menu.