Appendix A: Specifications
154
TDS1000/2000-Series Digital Oscilloscope User Manual
Oscilloscope Specifications (Cont.)
Vertical
Analog Bandwidth in
Sample and Average
modes at BNC or with
P
b
D
C
TDS1002 and
TDS2002
TDS1012,
TDS2012 and
TDS2014
TDS2022 and TDS2024
P2200 probe, DC
Coupled
60 MHz{* 100 MHz{* 200 MHz{*
32° F - 104° F(0° Cto+40° C)
160 MHz{*
32° F - 122° F(0° Cto+50° C)
20 MHz* (when vertical scale is set to < 5 mV)
Analog Bandwidth in
Peak Detect mode
(
/
d
i
/
d
i
*
*
)
TDS1002 and
TDS2002
TDS1012, TDS2012, TDS2014, TDS2022 and
TDS2024
(50 s/div to 5 s/div**),
t
y
p
i
c
a
l
50 MHz{* 75 MHz{*
t
y
p
i
c
a
l
20 MHz* (when vertical scale is set to < 5 mV)
Selectable Analog
Bandwidth Limit,
typical
20 MHz*
Lower Frequency
Limit, AC Coupled
10 Hz at BNC
1 Hz when using a 10X passive probe
Rise Time at BNC,
typical
TDS1002 and
TDS2002
TDS1012,
TDS2012 and
TDS2014
TDS2022 and TDS2024
<5.8ns <3.5ns <2.1ns
Peak Detect
Response**
Captures 50% or greater amplitude of pulses 12 ns wide typical
(50 s/div to 5 s/div) in the center 8 vertical divisions
{ When vertical scale is set to 5mV.
* Bandwidth reduced to 6 MHz with a 1X probe.
** The oscilloscope reverts to Sample mode when the SEC/DIV (horizontal scale)
is set from 2.5 s/div to 5 ns/div on 1 GS/s models, or from 2.5 s/div to
2.5 ns/div on 2 GS/s models. The Sample mode can still capture 10 ns glitches.