The controls that roll the listing in all three menus, the normal State listing,
the Reference listing, and the Difference listing are synchronized unless the
number of pre-trigger states differ between the Reference listing and the
acquired data.
This means that when you change the current row position in the Difference
listing, the analyzer automatically updates the current row in the acquired
State listing, Reference listing and vice-versa.
If the three listings are synchronized and you re-acquire data, the Reference
listing may have a different number of pre-trigger states depending on the
state trace trigger criteria. The Reference listing can be resynchronized to
the State and Difference listings by entering the desired state (acquisition
memory) location from the pop-up keypad.
This allows you to view corresponding areas of all lists, to cross-check the
alignment, and to analyze the bits that do not match.
The Compare Menu
Difference Listing Field