Time Tags
If you select Count Time, the system places a time stamp on all displayed
data. Data stored before the trigger has negative values and data stored after
the trigger has positive values. Time tag numbering is set to be either
relative to the previous memory location or absolute from the trigger point.
Select Absolute or Relative by toggling the Absolute/Relative field. Time tag
resolution is 8 ns.
Memory Depth Considerations
If you have all pod pairs assigned, the state acquisition memory is reduced by
half when time or state tags are turned on. To retain the full memory depth
when using time or state tags, you must have at least one unassigned pod
pair.
The Trigger Menu
Count Field (State and State Compare only)