Teledyne Model 300EU Telescope User Manual


 
TROUBLESHOOTING & REPAIR M300EU
Addendum to M300E/EM Manual P/N 04288
11.5. SUBSYSTEM CHECHOUT
11.5.1. RELAY BOARD
The information found in Section 11.5.5 of the M300E/EM Operators Manual (P/N 04288) is applicable to the
M300EU with the following exception:
The following table replaces Table 11-8 of the M300E/EM Operators Manual.
Table 11-4: Relay Board Control Devices
FUNCTION
CONTROL
DEVICE
IN SOCKET
Wheel Heater K1 Yes
Bench Heater K2 Yes
OVEN HEATER K3 YES
IZS Valves U4 Yes
IR Source Drive U5 No
11.5.2. MOTHERBOARD
11.5.2.1. A/D Functions
The M300EU makes higher resolution CO measurements than the M300E/EM means that it is susceptible to
temperature coefficient issues related to the motherboard’s A-to-D converter that could occur if the temperature
of the motherboard differs significantly from that of the rest of the analyzer’s oven area.
Compare the values of the BOX TEMP & OVEN TEMP test functions. If they are significantly different
check the function of the:
Oven theaters.
The oven heater fans.
The motherboard stabilization fan.
The relays controlling the oven heaters.
No abnormal blockages to airflow inside the oven area exist.
66 05514 Rev A1