CALIBRATION PROCEDURES M300EU
Addendum to M300E/EM Manual P/N 04288
MANUAL CALIBRATION WITH ZERO/SPAN VALVES
The information found in Section 7.4 of the M300E/EM Operators Manual (P/N 04288) is applicable to the
M300EU with the following exception(s).
• Set the display to show the STABILITY test function.
• There is no GAS TO CAL step.
MANUAL CALIBRATION CHECKS WITH ZERO/SPAN VALVES
The information found in Section 7.5 of the M300E/EM Operators Manual (P/N 04288) is applicable to the
M300EU with the following exception(s).
• Set the display to show the STABILITY test function.
• There is no GAS TO CAL step.
CALIBRATION WITH REMOTE CONTACT CLOSURES
The information found in Section 7.6 of the M300E/EM Operators Manual (P/N 04288) is applicable to the
M300EU with the following exception(s).
NOTE
It is required that the analyzer be forced to perform an auto-reference measurement just prior to
calibration, however this cannot be forced via the analyzer’s digital control inputs.
The A-REF measurement can only be initiated using either the instrument’s front pane interface or via
Teledyne Instruments’ APICOM emulator software.
AUTOMATIC ZERO/SPAN CAL/CHECK (AUTOCAL)
The information found in Section 7.7 of the M300E/EM Operators Manual (P/N 04288) is applicable to the
M300EU with the following exception(s).
• The following table replaces Table 7-2 of the M300E/EM Operators Manual (P/N 04288).
Table 7-1: AUTOCAL Modes
MODE NAME ACTION
DISABLED
Disables the Sequence
ZERO
Causes the Sequence to perform a zero calibration/check
ZERO-SPAN Causes the Sequence to perform a zero and span concentration
calibration/check
SPAN Causes the Sequence to perform a span concentration
calibration/check
AUTO-REF
Causes the analyzer to perform an auto-reference measurement.
CO CALIBRATION QUALITY
The information found in Section 7.8 of the M300E/EM Operators Manual (P/N 04288) is applicable to the
M300EU.
CALIBRATION OF OPTIONAL SENSORS
The following Sections do not apply to the M300EU:
• 7.9.1 SECTION O
2
SENSOR CALIBRATION PROCEDURE
• 7.9.2 CO
2
SENSOR CALIBRATION PROCEDURE
42 05831 Rev. 01