Teledyne Model 300EU Telescope User Manual


 
THEORY OF OPERATION M300EU
Addendum to M300E/EM Manual P/N 04288
10.3.2. THE RELAY PCA
The information found in Section 10.4.5 of the M300E/EM Operators Manual (P/N 04288) is applicable to the
M300EU with the following exception(s):
10.3.2.1. Temperature Control of the Convection Oven
The information found in Section 10.4.5.1 of the M300E/EM Operators Manual (P/N 04288) is applicable to the
M300EU with the following exception(s):
There are two heaters attached to either side of the optical bench (see Figure 3-1) that radiate heat into
the insulated, oven area of the analyzer. Each heater has a fan attached to it that circulates the heated
air throughout the oven airspace.
The CPU senses the temperature level form different places inside the oven area and determines the
overall temperature of the oven. Eventually, all of the components within the oven area will stabilize at
the same temperature. This usually takes about 2 hours after the instrument is turned ON.
A separate fan blows directly on the A-to-D conversion circuitry of the motherboard to more keep it at
the same temperature as the oven airspace in order to eliminate any temperature-coefficient related
offset.
Figure 10-5: M300EU Heating Control Block Diagram
54 05831 Rev. 01