M300EU OPERATING INSTRUCTIONS M300EU
Addendum to M300E/EM Manual P/N 04288
6.3.4. RNGE Æ MODE Æ DUAL: CONFIGURING THE M300EU ANALYZER
FOR DUAL RANGE MODE
DUAL range mode allows the A1 and A2 outputs to be configured with separate reporting range spans as well
as separate electronic signal levels. The analyzer software calls these two ranges LOW and HI.
• The LOW range setting corresponds with the analog output labeled A1 on the rear panel of the
instrument and is viewable via the test function RANGE 1.
• The HIGH range setting corresponds with the A2 output and is viewable via the test function RANGE 2.
• While the software labels these two ranges as LOW and HI, when in DUAL mode their upper limits need
not conform to that convention. The upper span limit of the LOW/RANGE1 can be a higher number than
that of HI/RANGE 2
To set the ranges press following keystroke sequence:
.
SETUP X.X PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK MORE EXIT
SAMPLE RANGE=50.000 PPM CO= XXXX
<TST TST> CAL SETUP
SETUP X.X RANGE MODE MENU
MODE SET UNIT EXIT
SETUP X.X RANGE MODE:SNGL
SNGL DUAL AUTO ENTR EXIT
SETUP X.X RANGE MODE:SNGL
SNGL DUAL AUTO ENTR EXIT
SETUP X.X RANGE MODE MENU
MODE SET UNIT EXIT
SETUP X.X HIGH RANGE:500.0 Conc
00500.0ENTREXIT
Toggled these keys to
select the upper
SPAN limit for the
reporting range
EXIT discards the new
setting
ENTR accepts the
new setting
SETUP X.X LOW RANGE:500.0 Conc
00500.0ENTREXIT
The LOW and HIGH
ranges have separate
slopes and offsets for
computing the CO
concentration.
The two ranges must
be independently
calibrated
.
EXIT discards the
new setting
ENTR accepts the
new setting