Teledyne Model 300EU Telescope User Manual


 
M300EU TROUBLESHOOTING & REPAIR
(Addendum to M300EU Manual - P/N 04145)
Table 11-2: Test Functions - Indicated Failures
TEST
FUNCTIONS
(As Displayed)
INDICATED FAILURE(S)
PHT DRIVE
If this drive voltage is out of range it may indicate one of several problems:
- A poor mechanical connection between the various components in inside the detector housing
- An electronic failure of the IR Photo-Detector’s built-in cooling circuitry, or;
- A temperature problem inside the analyzer chassis. In this case other temperature warnings would also be
active such as OVEN TEMP WARNING, BENCH TEMP WARNING or BOX TEMP WARNING.
SLOPE
Values outside range indicate
Contamination of the zero air or span gas supply
Instrument is Miscalibrated
Blocked gas flow
Contaminated or leaking GFC wheel (either chamber)
Faulty IR photo-detector
Faulty sample faulty IR photo-detector pressure sensor (P1) or circuitry
Invalid M/R ratio (see above)
Bad/incorrect span gas concentration due.
OFFSET
Values outside range indicate
Contamination of the zero air supply
Contaminated or leaking GFC wheel (either chamber)
Faulty IR photo-detector
05514 Rev A1 63