M300EU OPERATING INSTRUCTIONS M300EU
Addendum to M300E/EM Manual P/N 04288
6.7. ANALOG OUTPUT CONFIGURATION
6.7.1. ANALOG I/O CONFIGURATION
Because the manner in which the analog outputs are implemented differs between the M300E/EM and the
M300EU, there are significant differences in the manner in which the ANALOG I/O are manipulated.
THE ANALOG I/O CONFIGURATION SUBMENU
This section replaces Section 6.13.5.1 of the M300E/EM Operators Manual (P/N 04288).
Table 6-5: DIAG - Analog I/O Functions
SUB MENU FUNCTION
AOUT CALIBRATED
Initiates a calibration of the A1, A2 and A4 analog output channels that
determines the slope and offset inherent in the circuitry of each output.
These values are stored in the and applied to the output signals by the CPU
automatically
CONC_OUT_1
1
Sets the basic electronic configuration of the A1 output. There are four
options:
• RANGE: Selects the signal type (voltage or current loop) and level of the
output
• A1 OFS: Allows them input of a DC offset to let the user manually adjust
the output level
• AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature
• CALIBRATED: Performs the same calibration as AOUT CALIBRATED,
but on this one channel only.
CONC_OUT_2
1
Sets the basic electronic configuration of the A2 output. There are three
options:
• RANGE: Selects the signal type (voltage or current loop) and level of the
output
• A2 OFS: Allows them input of a DC offset to let the user manually adjust
the output level
• AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature
• CALIBRATED: Performs the same calibration as AOUT CALIBRATED,
but on this one channel only.
TEST OUTPUT
1
Sets the basic electronic configuration of the A4 output. There are three
options:
• RANGE: Selects the signal type (voltage or current loop) and level of the
output
• A4 OFS: Allows them input of a DC offset to let the user manually adjust
the output level
• AUTO CAL: Enables / Disables the AOUT CALIBRATION Feature
• CALIBRATED: Performs the same calibration as AOUT CALIBRATED,
but on this one channel only.
AIN CALIBRATED
Initiates a calibration of the A-to-D Converter circuit located on the Mother
Board.
1
Changes to RANGE or REC_OFS require recalibration of this output.
34 05831 Rev. 01