Using Additional Capabilities 6 - 5
•
t2 from TRIG Measures the time difference between
the trigger occurrence and the right cursor.
•
V1 Measures the voltage amplitude where the left
cursor crosses the waveform.
•
V2 Measures the voltage amplitude where the right
cursor crosses the waveform.
•
Vrms Measures the Root Mean Square (rms) voltage
between the cursors.
•
V DC (mean) Measures the MEAN (average of read-
ings) value of waveform between the cursors.
•
Vpeak/peak Measures the PEAK to PEAK value of the
waveform between the cursors.
•
Vmax peak Measures the maximum peak value of the
waveform between the cursors.
•
Vmin peak Measures the minimum peak value of the
waveform between the cursors.
•
frequency Measures the frequency of the waveform
between the cursors. At least 1
1
/
2
cycles of this wave-
form must occur between the cursors. You can scroll a
waveform through the cursors by pressing .
Read the frequency change in the bottom display.
•
rise time Measures the rise time between 10% and
90% points of the waveform. The 0% and 100% points
are determined by the intersection of the waveform and
the cursors. Set the rising slope to at least four time
divisions for a precise measurement.
•
phase Measures the phase shift between two
waveforms. You can select upto three waveforms for
phase measurements to a reference waveform as
selected with Measure on Waveform.