Tektronix AWG520 Portable Generator User Manual


 
Performance Tests
AWG510 & AWG520 Service Manual
4-19
AWG510 & AWG520 Test Record (Page 1 of 4)
Serial Number: Certificate Number:
Calibration Date:
Technician:
Clock frequency tests Low limit Test result High limit
10 MHz clock frequency accuracy 9,999,990 Hz 10,000,010 Hz
Amplitude and offset tests Low limit Test result High limit
CH1
Amplit de acc rac
20 mV 17.7 mV 22.3 mV
A
mp
lit
u
d
e accuracy
200 mV 195 mV 205 mV
2V 1.968 V 2.032 V
CH2 or CH1
Amplit de acc rac
20 mV 17.7 mV 22.3 mV
A
mp
lit
u
d
e accuracy
200 mV 195 mV 205 mV
2V 1.968 V 2.032 V
CH1
Offset acc rac
0V -10 mV +10 mV
Off
se
t
accuracy
1V 0.980 V 1.020 V
-1 V -1.020 V -0.980 V
CH2 or CH1
Offset acc rac
0V -10 mV +10 mV
Off
se
t
accuracy
1V 0.980 V 1.020 V
-1 V -1.020 V -0.980 V
Direct DA Output tests Low limit Test result High limit
CH1 Amplitude accuracy 0.45 V 0.55 V
CH2 or CH1 Amplitude accuracy 0.45 V 0.55 V
CH1 Offset accuracy -0.297 V -0.243 V
CH2 or CH1 Offset accuracy -0.297 V -0.243 V
CH1 Pulse rise time 700 ps
CH2 or CH1 Pulse rise time 700 ps