Tektronix AWG520 Portable Generator User Manual


 
Performance Verification
AWG510 & AWG520 Service Manual
4-7
Table 4-1: Diagnostic categories and error codes
Categories Error codes Descriptions
System 2100
2101
2102
2103
2104
2105
2106
Bios test error
Realtime clock power fail
CMOS checksum fail
FDD missing fail
Memory size fail
Flash disk fail
Date and time fail
2110
2111
2112
2113
2114
2115
2116
Front panel test error
Configuration fail
Communication fail
RAM fail
ROM fail
AD fail
Timer fail
2300
2301
A30 board test error
Unrecognized A30 board
2400
2401
2402
Clock delay data test error
Not found
Checksum fail
2700
2701
2702
Cal data test error
Uncaled
Checksum fail
Run mode 3100
3101 to 3104
CTRL1 registers test error
BIT0 to BIT3 fail (BIT0: 3101)
3200
3201 to 3216
Event table memory data bus error
DATA00 to DATA15 fail (DATA00: 3201)
3250
3251 to 3274
Event table address bus error
ADD00 to ADD23 fail (ADD00: 3251)
3300
3301 to 3302
Event table CS test error
CS0 to CS1 fail (CS0: 3301)
3350
3351 to 3352
Event memory cell test error
CHIP0 to CHIP1 fail (CHIP0: 3351)
Clock 4100
4101
4102
4103
4104
A40 clock board test error
PLL unlock check at 500 MHz faile
PLL lock check at 630 MHz fail
PLL lock check at 1350 MHz fail
PLL unlock check at 1500 MHz fail