Tektronix AWG520 Portable Generator User Manual


 
Performance Verification
AWG510 & AWG520 Service Manual
4-49
H Verify that the CH1
OUTPUT is off.
If the CH1
LED is on, push CH1 OUTPUT (front-panel) to turn the
LED off.
4. Set the trigger level to 5 V by following the substeps below:
a. Set the trigger level.
H Push SETUP (front-panel)!Trigger (bottom)!Level (side).
H Push 5 and ENTER keys in this order.
b. Set the offset level of generator.
H Push generator output ON.
H Push Cursor, %, &,
, keys as the high level of a pulse to be set
to 4.65V.
c. Verify that no waveform is displayed on the oscilloscope.
Trigger Signal CH2
(4.65 V level)
NOTE: In step 4b this
voltage level equals
4.65 V. In step 4d this
voltage level equals
5.35 V.
Figure 4-17: Trigger Signal (+5V check1)
d. Push Cursor, %, &,
, keys as the high level of a pulse to be set to
5.35V.
e. Verify that a sine wave is displayed on the oscilloscope.