Slope . ..................... Positive
Level ...................... 0V
Mode . ..................... Auto
2. Set the AWG500–Series Waveform Generator controls and select the
waveform file:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)!
System (bottom)!Factory Reset (side)!OK (side).
b. Select the waveform for the CH1: Load the MODE.WFM as referring to
the procedures on page 4–15.
c. Select the waveform for the CH2 (for the AWG520 only):
H Push the CH2 button.
H Select the waveform: Load the MODE.WFM as referring to the
procedures on page 4–15.
d. Set the clock frequency:
H Push HORIZONTAL MENU (front–panel)!Clock (side).
H Enter numeric value of 100M: Push 1, 0, 0 and M (SHIFT+7) keys
in this order.
e. Set the marker controls for the CH1:
H Push the CH1 button.
H Push VERTICAL MENU (front–panel)!Marker... (side)!
Marker1 Low Level (side).
H Enter numeric value of –2: Push –, 2 and ENTER keys in this order.
H Push the Marker2 Low Level side button.
H Enter numeric value of –2: Push –, 2 and ENTER keys in this order.
f. Set the marker controls for the CH2: Push CH2 button instead of CH1
to do the procedures in step 2.e.
3. Turn on the AWG500–Series Waveform Generator output: Push the RUN ,
CH1 OUT and CH2 OUT button so that the RUN LED, CH1 OUT LED
and CH2 OUT LED light.