Tektronix AWG520 Portable Generator User Manual


 
Performance Verification
4-76
AWG510 & AWG520 Service Manual
Clock Output Tests
These procedures check the AWG500–Series Waveform Generator clock output
signal.
Equipment
required
A 50ĂBNC coaxial cable, and an oscilloscope.
Prerequisites The instrument must meet the prerequisites listed on page 4-13.
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG500–Series Waveform
Generator CLOCK OUT output through a 50 BNC coaxial cable to
the oscilloscope input connector (see Figure 4–36).
Connect the BNC
connector end to
CLOCK OUT connector
Oscilloscope
AWG500 Series Waveform
Generator rear panel
50 W coaxial cable
Figure 4-36: Clock output initial test hookup
b. Set the oscilloscope controls:
Vertical . ........................ CH1
CH1 coupling . ............... DC
CH1 scale . ................. 0.5V/div
CH1offset .................. -1.3 V
CH1 input impedance . ......... 1MW
Horizontal
Sweep ..................... 5ns/div
Trigger
Source . .................... CH1
Coupling .................... AC
Slope ...................... Positive