Tektronix AWG520 Portable Generator User Manual


 
Performance Verification
AWG510 & AWG520 Service Manual
4-77
Level . ..................... 0V
Mode . ..................... Auto
2. Set the AWG500–Series Waveform Generator controls:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)!
System (bottom)!Factory Reset (side)!OK (side).
b. Select the waveform file: Load the MODE.WFM as referring to the
procedures on page 4–15.
c. Set the clock frequency:
H Push HORIZONTAL MENU (front–panel)!Clock (side).
H Enter numeric value of 50M: Push 5, 0 and M (SHIFT+7) keys in
this order.
3. Turn on the AWG500–Series Waveform Generator output: Push the RUN
button so that the LEDs above the RUN button lights.
4. Check the clock output signal: Check that the clock signal amplitude is equal
to or larger than 0.5 V, and the clock signal period is 20 ns.
5. End procedure: Disconnect the oscilloscope.