Chapter 5: Troubleshooting
Troubleshooting the Agilent 1690A,AD-series
Possible problem with
host operating system.
Consult host PC
documentation.
Is the IEEE 1394
cable connected and
seated?
Perform " To
verify connectivity"
in Chapter 5.
Remove the cover and
observe the status LEDs
on the acquisition board.
Is connectivity
verified?
Uninstall, then reinstall
the Agilent Logic
Analyzer application.
Uninstall, then reinstall
the Agilent Logic
Analyzer application.
Does the
application
software launch?
Do error
messages appear?
No
Yes
No
Yes
Reconnect or reseat
IEEE 1394 cable.
Possible problem with
acquisition board.
Replace acquisition bd.
Are any green
LEDs blinking?
Possible problem with
PCI IEEE 1394 board on
host PC. Replace
the board.
Yes
No
Yes
No
Yes
No
3A
1
3B
1