Agilent Technologies 1690 Weather Radio User Manual


 
64
Chapter 3: Testing Performance
To test the single-clock, multiple-edge, state acquisition
to turn off the other clocks.
d Connect the clock to be tested to the pulse generator channel 1 output.
e Click OK to close the Analyzer Setup dialog.
4 Verify the test data:
a Click the Run icon.
b If you have not already done so, do “Set up the Markers:” on page 34.
c If the "can't find 4096 occurence(s)" does not appear, then the test
passes.
The test passes when the logic analyzer finds all occurrences of the patterns
programmed into the Markers. If the test passes, record a "Pass" in the
performance test record under single-clock single-edge next to the clock and
edge being tested.
5 Test the next clock:
a Click the Sampling Setup icon.
b Disconnect the clock just tested from the pulse generator.
c Repeat steps 3, 4, and 5 for the next clock configuration listed in step 4
until all listed clock combinations have been tested.
6 Test the next setup/hold combination:
a Click the Bus/Signal Setup icon.
b Disconnect the clock just tested from the pulse generator.
c Repeat steps 1 through 6 for the next setup/hold combination listed in
step 1 in page 61.