116
Chapter 6: Replacing Assemblies
1680A,AD-series disassembly/assembly
6 Slide the acquisition board out the rear panel of the logic analyzer.
Steps to remove probe shroud from acquisition board.
7 Using a hex screwdriver, remove two hex nuts from the acquisition board
trigger BNC connectors.
8 Using a Torx T-10 screwdriver, remove five screws that secure the probe
shroud to the acquisition board.
Reverse this procedure to install the acquisition board.
If the probe shroud requires replacing, then the probe shroud label (part number
01680-94312) must also be ordered and installed on the replacement probe
shroud.