Command Reference 9-127
*PUDPM6680B/81/85
8 <Arbitrary block program data>
Protected User Data
Protected user data. This is a data area in which the user may write any data up to
64 characters. The data can always be read, but you can only write data after un
-
protecting the data area. A typical use would be to hold calibration information, us
-
age time, inventory control numbers, etc.
The content at delivery is: #234 FACTORY CALIBRATED ON: 19YY-MM-DD
–
YY = year, MM = month, DD = day
Returned format: <Arbitrary block response data>¿
–
Where:
<arbitrary block program data> is the data last programmed with *PUD.
Example
Send ® :SYST:UNPR; *PUD 8 #240Calibrated 8 1993-07-16, 8 inven
-
tory 8 No.1234
# means that <arbitrary block program data> will follow.
2 means that the two following digits will specify the length of the data block.
40 is the number of characters in this example.
Complies to standards: IEEE 488.2 1987.
*RCLPM6680B/81/85
8 <Decimal data>
Recall
Recalls one of the up to 20 previously stored complete instrument settings from the
internal nonvolatile memory of the instrument.
Memory number 0 contains the power-off settings for PM6685. For PM6681 mem
-
ory number 0 contains the power-off settings until PRESET is pressed. After pre
-
set, memory 0 contains the pre-preset settings.
Parameters:
<Decimal data> = a number between 0 and 19.
Example:
SEND® *RCL 8 10¿
Complies to standards: IEEE 488.2 1987.