9-126 Command Reference
*PMC PM6680B/81/85
Purge Macros
Removes all macro definitions.
Example: *PMC
See also:
:MEMory:DELete:MACRo 8 ‘<Macro-name>’ if you want to remove a single
macro.
Complies to standards: IEEE 488.2 1987.
*PSC PM6680B/81/85
8 <Decimal data>
Power-on Status Clear
Enables/disables automatic power-on clearing. The status registers listed below
are cleared when the power-on status clear flag is 1. Power-on does not affect the
registers when the flag is 0.
Service request enable register (*SRE)
Event status enable register (*ESE)
Operation status enable register (:STAT:OPER:ENAB)
Questionable data/signal enable register (:STAT:QUES:ENAB)
Device enable registers (:STAT:DREG0:ENAB)
*RST does not affect this power-on status clear flag.
Parameters: <Decimal data> = a number that rounds to 0 turns off automatic power-on
clearing. Any other value turns it on.
Returned Format: «1|0»¿
1 is enabled and 0 is disabled.
Example: *PSC 8 1
This example enables automatic power-on clearing.
Complies to standards: IEEE 488.2 1987.