Fluke PM6685 Outdoor Timer User Manual


 
9-6 Command Reference
:ARM :COUNt PM6680B/81/85
8«<Numeric value>|MIN|MAX»
No. of Measurements on each Bus arm
This count variable controls the upward exit of the “wait-for-bus-arm” state
(:ARM:STARt:LAY1). The counter loops the trigger subsystem downwards
COUNt number of times before it exits to the idle state.
This means that a COUNt No. of measurements can be done for each Bus arming
or INITiate.
L
The actual number of measurements made on each INIT is equal to:
(:ARM:START:COUNT)*(:TRIG:START:COUNT)
Parameters:
<Numeric value> is a number between 1 and 65 535. (1 switches the function OFF.)
MIN gives 1
MAX gives 65 535
Returned format: <Numeric value>¿
Example:
SEND® :ARM:COUN8100¿
*RST condition: 1
Complies to standards: SCPI 1991.0, confirmed