Fluke PM6685 Outdoor Timer User Manual


 
9-96 Command Reference
:ROSCillator :SOURce PM6680B/81/85
8 «INT|EXT»
Select Reference Oscillator
Selects the signal from the external reference input as timebase instead of the in
-
ternal timebase oscillator.
Returned format: <INT|EXT>¿
*RST condition: INT
Complies to standards: SCPI 1991.0, confirmed.
:SDELay PM6685
8 «<Numeric value>|MIN|MAX»
BURST/PRF Synchronization Delay
Sets the synchronization delay time used in FREQuency:BURSt | PRF measure-
ments.
Parameter range: 200 ns to 1.6777215 s
*RST condition: 10 ms