Tektronix TDS 644A Sprinkler User Manual


 
Vertical Control
Reference
3Ć138
Bandwidth
To eliminate the higher frequency components, change the bandwidth of the
selected channel:
Press VERTICAL MENU Bandwidth (main) Full, 100 MHz,or20ĂMHz
(side).
Fine Scale
Press VERTICAL MENU Fine Scale (main) to make fine adjustments to
the vertical scale using the general purpose knob or the keypad.
Position
Press VERTICAL MENU Position (main) to let the general purpose knob
control the vertical position. Press Set to 0 divs (side) if you want to reset
the reference point of the selected waveform to the center of the display.
Offset
Offset lets you subtract DC bias from the waveform, so the oscilloscope can
acquire the exact part of the waveform you are interested in.
Offset is useful when you want to examine a waveform with a DC bias. For
example, you might be trying to look at a small ripple on a power supply
output. It may be a 100 mV ripple on top of a 15 V supply. Using offset, you
can display the ripple and scale it to meet your needs.
To use offset, press VERTICAL MENU Offset (main). Use the general
purpose knob to control the vertical offset. Press Set to 0 V (side) if you
want to reset the offset to zero.
See Acquisition, on page 2Ć19.
See Scaling and Positioning Waveforms, on page 2Ć22.
For More
Information