Tektronix TDS 350 Sprinkler User Manual


 
Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć33
ACQuire:STOPAfter
Tells the oscilloscope when to stop taking acquisitions. This is equivalent to
setting Stop After in the Acquire menu.
Acquisition
ACQuire:MODe, ACQuire:STATE
ACQuire:STOPAfter { RUNSTop | SEQuence}
ACQuire:STOPAfter?
ACQuire
:
STOPAfter
<Space>
?
RUNSTop
SEQuence
RUNSTop specifies that the run and stop state should be determined by the
user pressing the frontĆpanel RUN/STOP button.
SEQuence specifies single sequence" operation, where the oscilloscope
stops after it has acquired enough waveforms to satisfy the conditions of the
acquisition mode. For example, if the acquisition mode is set to sample, and
the horizontal scale is set to a speed that allows realĆtime operation, then the
oscilloscope stops after digitizing a waveform from a single trigger event.
However, if the acquisition mode is set to average 16 waveforms, then the
oscilloscope stops only after acquiring all 16 waveforms. The ACQuire:
STATE command and the frontĆpanel RUN/STOP button also stop acquisiĆ
tions when the oscilloscope is in single sequence mode.
NOTE
If you set the acquisition system to single sequence, envelope
mode, and set the number of envelopes to infinity, the oscilloscope
will envelope a maximum of 257 acquisitions.
ACQUIRE:STOPAFTER RUNSTop
sets the oscilloscope to stop acquisition when the user presses the
frontĆpanel RUN/STOP button.
ACQUIRE:STOPAFTER?
might return SEQUENCE.
Group:
Related Commands:
Syntax:
Arguments:
Examples: