H Assign each device on the bus a unique device address; no two devices
can share the same device address.
H Do not connect more than 15 devices to any one bus.
H Connect one device for every 6 feet (2 meters) of cable used.
H Do not use more than 65 feet (20 meters) of cable to connect devices to a
bus.
H At least twoĆthirds of the devices on the network must be turned on while
the network is operating.
H Connect the devices on the network in a star or linear configuration as
shown in Figure 1Ć7. Do not use loop or parallel configurations.
Appendix C: Interface Specifications, gives additional information on the GPIB
configuration of the oscilloscope.