Tektronix TDS 350 Sprinkler User Manual


 
Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć115
MEASUrement:IMMed:UNIts? (Query Only)
Returns the units for the immediate measurement.
Measurement
MEASUrement:IMMed:TYPe
MEASUrement:IMMed:UNIts?
:MEASUrement
IMMed
:
UNIts
?
<QString> returns ”V” for volts, ”s” for seconds, ”Hz” for hertz, or ”%” for
percent.
MEASUREMENT:IMMED:UNITS?
might return ”s”, indicating that the units for the immediate measureĆ
ment are seconds.
MEASUrement:IMMed:VALue? (Query Only)
Executes the immediate measurement specified by the MEASUreĆ
ment:IMMed:TYPe command. The measurement is taken on the source
specified by the MEASUrement:IMMed:SOUrce command.
Measurement
MEASUrement:IMMed:VALue?
:MEASUrement
IMMed
:
VALue
?
<NR3>
Group:
Related Commands:
Syntax:
Returns:
Examples:
Group:
Syntax:
Returns: