Tektronix TDS 350 Sprinkler User Manual


 
Command Descriptions
TDS 310, TDS 320 & TDS 350 Programmer Manual
2Ć133
RECAll:SETUp (No Query Form)
Restores a stored or factory frontĆpanel setup of the oscilloscope. This comĆ
mand is equivalent to selecting Recall Saved Setup or Recall Factory Setup
in the Save/Recall Setup menu. This command is the same as the FACtory
command.
Save and Recall
FACtory, *RCL, *RST, *SAV, SAVe:SETUp, TEKSecure
RECAll:SETUp { FACtory | <NR1> }
<Space>
RECAll
:
FACtory
SETUp
<NR1>
FACtory selects the factory setup.
<NR1> is a value in the range from 1 to 10 and specifies a setup storage
location. Using an outĆofĆrange value causes an execution error (222, Data
out of range").
RECALL:SETUP FACTORY
recalls the frontĆpanel setup to its factory defaults.
REM (No Query Form)
Specifies a comment. Comments are ignored by the oscilloscope.
Miscellaneous
REM <QString>
REM <Space>
<QString>
<QString> is a string that can have a maximum of 80 characters.
REM ”This is a comment”
is ignored by the oscilloscope.
Group:
Related Commands:
Syntax:
Arguments:
Examples:
Group:
Syntax:
Arguments:
Examples: