Agilent Technologies n3306a Portable Generator User Manual


 
Performance Test and Calibration - B
95
VOLTAGE Calibration Program
10 ! This program calibrates the voltage mode for load modules
20 ! N3302A, N3303A, N3304A, N3305A and N3306A.
30 !
40 ! last update 9/6/00
50 !
60 ! Edit the interface address on line 100 if different then 705
70 !
80 ASSIGN @Ld TO 705
90 !
100 OUTPUT @Ld;"*RST"
110 OUTPUT @Ld;"CAL:STAT ON"
120 OUTPUT @Ld;"FUNC VOLT"
130 OUTPUT @Ld;"VOLT:RANG MAX"
140 OUTPUT @Ld;"SENS:VOLT:RANG MAX"
150 OUTPUT @Ld;"INP:STAT ON"
160 !
170 ! ******* Calibrating high volt range *******
180 !
190 PRINT TABXY(10,8),"See figure B-3 for voltage calibration"
200 PRINT TABXY(10,10),"Connect power leads to input terminals"
210 PRINT TABXY(10,12),"Connect + sense to + input terminal"
220 PRINT TABXY(10,14),"Connect - sense to - input terminal"
230 PRINT TABXY(10,16),"Connect DVM to +/- sense terminals"
240 PRINT TABXY(10,18),"See Table B-2 Voltage Mode Voltage / Current Settings"
250 PRINT TABXY(10,20),"Set power source to voltage and current for model to be
calibrated"
260 PRINT TABXY(10,24),"BE SURE REMOTE BUTTON ON REAR PANEL IS IN REMOTE
POSITION"
270 DISP "Press CONT when ready to calibrate voltage "
280 PAUSE
290 CLEAR SCREEN
300 !
310 PRINT TABXY(9,15);"CALIBRATING HIGH VOLTAGE RANGE"
320 OUTPUT @Ld;"CAL:LEV P1"
330 WAIT 1
340 INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp1
350 !
360 OUTPUT @Ld;"CAL:DATA";Vp1
370 !
380 OUTPUT @Ld;"CAL:LEV P2"
390 WAIT 1
400 INPUT "Measure voltage at +/- sense terminals and enter value in volts",Vp2
410 !
420 OUTPUT @Ld;"CAL:DATA";Vp2
430 !
440 ! ******* Calibrating low voltage range *******
450 !
460 OUTPUT @Ld;"VOLT:RANG MIN"
470 OUTPUT @Ld;"SENS:VOLT:RANG MIN"
480 !
490 CLEAR SCREEN
500 PRINT TABXY(9,15);"CALIBRATING LOW VOLTAGE RANGE"
510 !
520 OUTPUT @Ld;"CAL:LEV P1"
530 WAIT 1