Agilent Technologies n3306a Portable Generator User Manual


 
Front Panel Operation - 5
67
Figure 5-7. Toggled Transient Operation
Action Display
1. On the Function keypad, press Current. On the Entry keypad, press 5 Enter . CURR 5
2. On the Function keypad, press Current. On the Function keypad, press key until
transient current is displayed.
C:TLEV
3.
On the Entry keypad press 10 Enter.
C:TLEV 10
6. On the Function keypad, press Tran. On the Function keypad, press key until
transient mode is displayed.
T:MODE
7. To select the pulse mode, press the Input or Input key until the Toggle mode
is displayed. Then press Enter.
T:MODE TOGG
8. On the Function keypad, press the Tran key and select ON to activate the transient
mode.
TRAN ON
9. On the Function keypad, press Trigger Control, then press Enter. This initiates
the trigger system.
INIT:IMMED
10
Press Shift and Trigger. This switches to the toggled level. Each time a trigger is
sent, the input switches to the alternate level.
NOTE: The difference between programming a toggled level (C:TLEV) and a triggered level (C:TRIG),
is that a toggled level allows you to switch back and forth between the main (or immediate)
level and the toggled level. A triggered level, on the other hand, becomes the main level once it
is triggered. Subsequent triggers have no effect on the input level.
4 - Programming Lists
Lists are the most flexible means of generating complex input sequences. The following figure shows a current
input sequence generated from a list. The input current shown represents three different current pulses (8 A for 2
seconds, 6 A for 5 seconds, and 4 A for 7 seconds) separated by 3-second, 0-ampere intervals.
The list specifies the pulses as three current points (point 0, 2, and 4), each with its corresponding dwell point.
The intervals are three zero-current points (point 1, 3, and 5) of equal intervals. The count parameter causes the
list to execute twice when started by a single trigger.
NOTE: From the Recall Menu, execute the *RST command to reset the load module. This is necessary
because any previously programmed functions remain in effect until cleared.
Action Display
1. Press Current to access the Current Menu. Then press to access the current mode
command.
C:MODE FIXED
2.
On the Entry keypad, press Input or Input to scroll through the mode parameters
to obtain LIST and press Enter.
C:MODE LIST
3.
Access the List Menu by pressing Shift List. Press to scroll to the count command.
From the Entry keypad, change the list count from the default (1) to 2. Press Enter.
LST:CNT 2