Tektronix TDS3000 Sprinkler User Manual


 
TDS3000 Series Service Manual
1–1
Specifications
This appendix contains specifications for the TDS3000 series oscilloscopes. All
specifications are guaranteed unless noted as “typical.” Typical specifications are
provided for your convenience but are not guaranteed. Specifications that are
marked with the symbol are checked in Performance Verification.
All specifications apply to all TDS3000 models unless noted otherwise. To meet
specifications, two conditions must first be met:
The oscilloscope must have been operating continuously for ten minutes
within the operating temperature range specified.
You must perform the Compensate Signal Path operation described on page
2–21. If the operating temperature changes by more than 10° C, you must
perform the Compensate Signal Path operation again.
Table 1–1: Specifications
Acquisition
Acquisition modes Sample (Normal), Peak detect, Envelope, and Average
S
ingl
e
Seq
u
e
nc
e
Acquisition mode Acquisition stops after
Sample, Peak Detect One acquisition, all channels
simultaneously
Average, Envelope N acquisitions, all channels
simultaneously, N is settable from
2 to 256 (or for Envelope)