Signal Generator Specifications
286
Output
RF IN/OUT Connector:
Standard:
Level Range: −137 to −19 dBm into 50Ω
Level Accuracy:
±1.2 dB (level ≥ −127 dBm) (typical ±1.0 dB for all levels)
Reverse Power:
60 W continuous
100 W for 10 seconds/minute
Option 007:
Level Range: −137 to −5 dBm into 50 Ω]
Reverse Power:
2.4 W continuous
4.0 W for 10 seconds/minute
DUPLEX OUT Connector:
Level Range: −127 to +7 dBm into 50 Ω
Level Accuracy: ±1.0 dB
Reverse Power: 200 mW maximum
SWR:
RF IN/OUT: <1.5:1
DUPLEX OUT: <2.0:1 (level <−4 dBm)
Resolution: 0.1 dB