Agilent Technologies HP8648A Weather Radio User Manual


 
Using the Tracking Generator
219
Measurement Procedure:
On the Test Set:
1. Insert the System Support Test Card (HP 11807A Option 100) into the
Memory Card Slot.
2. Press the PRESET key.
3. Press the TESTS key.
Using the knob and data entry keys:
4. Select Card from the Location field.
5. Select CABLE_FLT from the Procedure field.
6. Select Edit Parm from the Test Function field.
7. Set cable length units to feet (0.000000) or meters (1.000000).
8. Enter cable type.
9. Enter length of cable under test (in feet or meters). Set this value to
approximately 1.5 times the estimated maximum length.
Test Set Options Required
Spectrum Analyzer/Tracking Generator (option 102).
Additional Equipment
Required
HP 11807A Option 100, a Power Splitter/Combiner (HP
11636A), and a 50 Load.
Special Test Considerations
Cable fault must be within 1000 feet of the Test Set.