Tektronix AFG3000 Portable Generator User Manual


 
Reference
2-10 AFG3000 Series Arbitrary/Function Generators Reference Manual
When sweep is taking place, the frequency sweeps from start frequency to stop fre-
quency. Refer to Figure 2-2.
Figure 2-2: Frequency sweep
The instrument sweeps from a low frequency to a high frequency when start fre-
quency is set to be lower than stop frequency (start frequency < stop frequency).
The instrument sweeps from a high frequency to a low frequency when start fre-
quency is set to be higher than stop frequency.
The frequency range differs based on the selected waveform types. For frequency
sweep, you can select a sine, square, ramp, or arbitrary waveform. Pulse, DC, and
Noise waveforms cannot be selected.
You can select one of the following two operation modes:
Repeat. A continuous sweep is output at a rate specified by the sweep time, hold
time, and return time.
Trigger. One sweep waveform is output when a trigger input is received.
You can select a trigger source from an external signal applied to the front-panel
Trigger Input connector, the trigger signal generated internally, the manual trigger,
or the remote interface.
Hold timeSweep time Return time
Stop freq.
Start freq.
Frequency
Frequency
span
Center
frequency
Time