Tektronix AFG3000 Portable Generator User Manual


 
Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual 3-79
[SOURce[1|2]]:SWEep:RTIMe
This command sets or queries the sweep return time. Return time represents the
amount of time from stop frequency through start frequency. Return time does not
include hold time.
Group
Source
Syntax
[SOURce[1|2]]:SWEep:RTIMe {<seconds>|MINimum|MAXimum}
[SOURce[1|2]]:SWEep:RTIMe? {MINimum|MAXimum}
Arguments
<seconds>::=<NRf>[<units>]
where
<NRf> is the return time in seconds.
<units>::=[ns | µs | ms | s]
Returns
<seconds>
Examples
SOURce1:SWEep:RTIMe 1ms
sets the CH 1 return time to 1 ms.
[SOURce[1|2]]:SWEep:SPACing
The [SOURce[1|2]]:SWEep:SPACing command selects linear or logarithmic
spacing for the sweep for the specified channel.
The query command returns the type for the sweep spacing for the specified
channel.
Group
Source
Syntax
[SOURce[1|2]]:SWEep:SPACing {LINear|LOGarithmic}
[SOURce[1|2]]:SWEep:SPACing?
Arguments
LINear sets the sweep spacing to linear.
LOGarithmic sets the sweep spacing to logarithmic.
Returns
LIN|LOG
Examples
SOURce1:SWEep:SPACing LINear
sets the CH1 sweep spacing to linear.