11.4 Confirma tion of Measurement Accuracy
It is extremely important to check operation and accuracy of the instrument using a supplied model eye.
We recommend you to check accuracy periodically.
When the measurement result of the model eye falls anywhere within the tolerance listed below,
measurement should be considered reliable and accurate. When the result exceeds the tolerance, contact
your dealer immediately.
Model Eye Data
SPH CYL R
Indicated value 0.25 0 0.25 Indicated value 0.03
Precise value of the supplied model eye is indicated on the model eye stand (VD =12).
Model eye
11.5 Periodica l Inspection and Maintenance
To prevent malfunction and accidents and maintain the performance and reliability of the product, it is
recommended to request your distributor for the periodical inspection and maintenance once a year.
The periodical inspection and maintenance include inspection of the function and performance of the
product, and cleaning, adjustment and replacement of consumable parts if necessary.
Note for setting of model eye
Remove a contact lens holder cap and set a model eye, checking it is not inclined in any
directions.
When the model eye is inclined, CYL value can not be correctly measured.
Set the model eye at the position where an alignment mark is located inside a reticle
mark and good focusing is achieved.
When conditions above have been satisfied, proceed to measurement.