Kowa RB-525-B04 Telescope User Manual


 
RB-525-B0434
10. Troubleshooting
If there are any malfunction found, refer to the table below to take appropriate measures.
CAUTION
Never disassemble, modify or repair the instrument.
Personal injury may result from electric shock.
Symptom Causes and Measures
The monitor and power indicator are not
turned on.
The power cord may be not properly connected. Make
sure to connect it securely.
Fuse may be blown. If so, replace it to the new one.
Fuse is blown when the power switch is
turned on.
Contact your local distributor immediately.
The monitor display suddenly disappears.
The saving function may be activated.
Press any switch to deactivate the saving function.
The main body can not be moved in a
horizontal direction.
Anti-sliding lock may be locked. Unscrew the lock
under the base.
The moving parts such as a joystick are not
moving appropriately.
Do not force to move the part. Contact your local
distributor or service person.
The apparatus does not print out.
Check the papers being set. Reload them if the papers
are out.
PRINT FORM in the MENU screen may be set for
OFF. Change the setting.
The printer papers come out but no printing.
The printer paper may be set in a wrong direction. Set
the paper properly.
The date setting becomes inaccurate.
The battery inside the apparatus may be dead. Keep the
power on for 24 hours to recharge it.
Contact your local distributor immediately if the situation does not improve even when the measures
mentioned above are taken.